School of Electrical and Computer Engineering, Georgia Institute of Technology, Atlanta, USA · 13 Jan 2020 · medicine

Discovery

Detecting Suspected Pump Thrombosis in Left Ventricular Assist Devices via Acoustic Analysis.

Source check: Source checked. Not independent scientific confirmation.

Finding: Among 13 HVAD recipients, reported accuracy was 88.9% and sensitivity 90.9%; post-thrombolysis recordings suggested abnormalities not identified by reference pump-power or biomarker measures. The small suspected-thrombosis series and post-treatment observations are not independent prospective validation or proof of clinical benefit.

Technology: Acoustic features from chest recordings were selected with LDH-based correlation analysis and used in logistic regression to detect suspected left-ventricular-assist-device pump thrombosis.

Publication: Peer-reviewed paper · Test context: retrospective or benchmark. Catalog categories, not an evidence grade.

AI contribution
Acoustic features from chest recordings were selected with LDH-based correlation analysis and used in logistic regression to detect suspected left-ventricular-assist-device pump thrombosis.
Test and result
Among 13 HVAD recipients, reported accuracy was 88.9% and sensitivity 90.9%; post-thrombolysis recordings suggested abnormalities not identified by reference pump-power or biomarker measures.
Limitation / still unknown
The small suspected-thrombosis series and post-treatment observations are not independent prospective validation or proof of clinical benefit.
Evidence sources and annotation context

Evidence context (descriptions follow the cited sources; test type is a catalog classification):

Evidence sources: https://pmc.ncbi.nlm.nih.gov/articles/PMC7380556/

Source checked · editorial details

The source page was checked against this entry. Independent confirmation is not assessed by this label.

Source type: paper. Paper: linked separately.

Review date: 22 Nov 2026. Review scope: Primary result-bearing material for the stated method, result and evaluation, with bibliographic version information..

No follow-up recorded; this is not a claim that no follow-up exists. Follow-up review scope and date are not recorded. No source-supported relationships recorded.

Editorial assessment: Impact 2/5 · Validation 2/5 · Novelty 2/5. Impact and novelty are editorial judgments; validation follows the annotated test type where one exists. Not an evidence grade. Scoring criteria.