Artificial intelligence to diagnosis distal radius fracture using biplane plain X-rays.
Source check: Source checked. Not independent scientific confirmation.
Finding / tested result: The study reports diagnostic accuracy on labeled X-ray patterns, with test performance around 89–94%.
Technology: A CNN diagnoses distal-radius fracture from biplane plain radiographs, a specific fracture-detection task.
Publication: Peer-reviewed paper · Test context: retrospective or benchmark. Catalog categories, not an evidence grade.
- AI contribution
- A CNN diagnoses distal-radius fracture from biplane plain radiographs, a specific fracture-detection task.
- Limitation / still unknown
- The diagnostic result is limited to labeled X-ray patterns and the tested dataset; it does not establish prospective clinical performance.
Evidence sources and annotation context
Evidence context (descriptions follow the cited sources; test type is a catalog classification):
Evidence sources: https://europepmc.org/article/MED/34823550 · https://pmc.ncbi.nlm.nih.gov/articles/PMC8620959/
Source checked · editorial details
The source page was checked against this entry. Independent confirmation is not assessed by this label.
Source type: paper. Paper: not linked in this catalog; this does not mean no paper exists.
Review date: 22 Nov 2026. Review scope: Primary result-bearing material for the stated method, result and evaluation, with bibliographic version information..
No follow-up recorded; this is not a claim that no follow-up exists. Follow-up review scope and date are not recorded. No source-supported relationships recorded.
Editorial assessment: Impact 3/5 · Validation 2/5 · Novelty 3/5. Impact and novelty are editorial judgments; validation follows the annotated test type where one exists. Not an evidence grade. Scoring criteria.