AI Science Tracker
EN

Multi-centre validation of an AI prostate MRI detection tool against radiologists

University College Hospital; University College London Hospitals NHS Foundation Trust; University College London · 28 Feb 2025 · medicine

How tested is it?

Checked on existing data

The level reflects the test the original source reports. This site does not rerun it, and it does not say whether the result is correct.

Discovery

Multi-centre validation of an AI prostate MRI detection tool against radiologists

Source check: Source checked. Not independent scientific confirmation.

Finding: A CE-marked deep-learning CAD device (Pi) trained on 793 patients was tested on a separate set of 252 patients across six sites and two manufacturers. Per-patient AUC was 0.91 versus 0.95 for MDT-supported radiologists, with AI non-inferior within a 10% margin. At its threshold AI had 95% sensitivity and 67% specificity, versus 99% and 73% for radiologists.

Technology: CE-marked deep-learning computer-aided detection device for Gleason grade group 2 or higher prostate cancer

Publication: Peer-reviewed paper · Test context: retrospective or benchmark. Catalog categories, not an evidence grade.

AI contribution
Device flags suspicious lesions.
Test and result
Separate multi-site retrospective validation set.
Limitation / still unknown
Retrospective; unbiopsied negatives assumed negative; commercial device; full text not reviewed.
Evidence sources and annotation context

Evidence context (descriptions follow the cited sources; test type is a catalog classification):

Evidence sources: https://www.ebi.ac.uk/europepmc/webservices/rest/search?query=DOI:%2210.1007/s00330-024-11323-0%22&resultType=core&format=json · https://api.openalex.org/works/doi:10.1007/s00330-024-11323-0

Source checked · editorial details

The source page was checked against this entry. Independent confirmation is not assessed by this label.

Source type: paper. Paper: not linked in this catalog; this does not mean no paper exists.

Review date: 26 Nov 2026. Review scope: Abstract read from Europe PMC metadata for this DOI. Author affiliations and publisher full text not accessed. Author affiliations from OpenAlex metadata.

No follow-up recorded; this is not a claim that no follow-up exists. Follow-up review scope and date are not recorded. No source-supported relationships recorded.

Editorial assessment: Impact 2/5 · Validation 2/5 · Novelty 2/5. Impact and novelty are editorial judgments; validation follows the annotated test type where one exists. Not an evidence grade. Scoring criteria.