Independent UK test of eight mammography AI algorithms as second reader and triage
Source check: Source checked (abstract only). Not independent scientific confirmation.
Finding: Eight commercial deep learning breast cancer detection algorithms from five vendors were evaluated retrospectively on 22,778 mammograms from two UK screening sites (Cambridge, Norwich; 2017) with 3-year follow-up. Standalone AUCs ranged 0.82-0.91, with one algorithm inferior to all others. Paired with the first human reader, all but one were noninferior in sensitivity to double reading (64.7-67.4% vs 67.9%) with superior specificity (97.5% vs 97.1%); low-suspicion triage ruled out 37.4-65.7% of mammograms while missing at most 1.0% of screen-detected cancers. Gains came with more arbitration, and the study is retrospective; prospective workflow effects were not measured.
Technology: Commercial deep learning breast cancer detection algorithms scoring digital screening mammograms.
Publication: Peer-reviewed paper · Test context: retrospective or benchmark. Catalog categories, not an evidence grade.
- AI contribution
- The deep learning algorithms produced the cancer suspicion scores whose standalone, paired-reader and triage performance is the finding.
- Test and result
- Retrospective evaluation on 22,778 UK screening mammograms with histopathology-confirmed screen-detected, interval and next-round cancers; noninferiority at a 5% margin.
- Limitation / still unknown
- Prospective performance, effects on reader behaviour and arbitration workload in practice, generalisability beyond two 2017 UK sites; vendor-level results and full methods not reviewed (full text not accessed).
Evidence sources and annotation context
Evidence context (descriptions follow the cited sources; test type is a catalog classification):
Evidence sources: https://www.ebi.ac.uk/europepmc/webservices/rest/search?query=DOI:%2210.1007/s00330-026-12997-4%22&resultType=core&format=json · https://api.openalex.org/works/doi:10.1007/s00330-026-12997-4
Source checked (abstract only) · editorial details
Only the abstract and bibliographic metadata were checked against this entry; the full text was not read. Independent confirmation is not assessed by this label.
Source type: paper. Paper: not linked in this catalog; this does not mean no paper exists.
Review date: 27 Nov 2026. Review scope: Independent review: structured abstract from Europe PMC, Crossref dates, OpenAlex affiliations. Springer full text not read.
No follow-up recorded; this is not a claim that no follow-up exists. Follow-up review scope and date are not recorded. No source-supported relationships recorded.
Editorial assessment: Impact 3/5 · Validation 2/5 · Novelty 2/5. Impact and novelty are editorial judgments; validation follows the annotated test type where one exists. Not an evidence grade. Scoring criteria.